Price: $ 5,710 |
| PRICES SUBJECT TO CHANGE
WITHOUT NOTICE |
 |
 |
Description |
 |
| The Model
460 is a 3-channel, high-resolution gaussmeter, for use with single
or multi-axis probes. It is capable of measuring magnetic fields from
0.001 Gauss to 300kGauss, with a DC accuracy +/- 0.10%. This unit
is designed for demanding applications, such as material analysis
and field mapping, where low-noise, high-resolution magnetic measurements
are necessary. Measurements of AC and DC magnetic fields are shown
in Gauss or Tesla on the two-line vacuum fluorescent display. The
Model 460 features RS232C and IEEE-488 ports, allowing the instrument
to be controlled via a remote computer or automation equipment. The
Model 460 can be used with a 3-axis probe or with three separate single-axis
probes, effectively providing three Model 450 gaussmeters in one instrument.
Probes are sold separately, and are available in a wide range of constructions. |
 |
| General
Specifications |
| Measurement
Ranges |
300mG
/ 3 G / 30 G / 300 G / 3 kG / 30 kG / 300 kG
(range is dependent on probe type) |
| Resolution |
0.001mG
(when used with filter on and appropriate probe) |
| Accuracy, DC (displayed) |
±
0.10% of reading; ± 0.005% of range |
| Accuracy, AC (displayed) |
±
2% of reading, rms (50 Hz - 60 Hz) |
| Frequency Range |
10 Hz
- 400 Hz |
| Update Rate |
4 readings
/ sec (display); up to 18 readings / sec on ports |
| Display Resolution |
±
5 3/4 digits (max.) |
| Measurement Modes |
DC,
RMS, Peak |
| Measurement Features |
Auto
Range, Max Hold, Relative Mode, Filter,
Vector Magnitude, Differential Reading |
| Measurement Units |
Gauss
or Tesla (with u, m, and k multipliers) |
| Probes |
Features
|
Linearity
correction, auto probe zero,
temperature correction |
| Types |
Multi-axis, axial,
transverse, flexible transverse |
| Compatibility |
Standard and custom
probes available |
| Connector |
15-pin D-style |
| Front Panel Display |
Vacuum-Flourescent,
high-visibiliy blue, 4 lines x 20 characters |
| Communication Ports |
RS232C
(300 / 1200 / 9600 Baud)
IEEE-488 (complies with IEEE-488.2) |
| Alarms |
Settings |
High/low set point,
inside/outside, audible |
| Actuators |
Display annunciator,
beeper |
| Temperature Range |
15 -
35° C (operating temp; 5 - 40° with reduced accuracy) |
| Input Power |
100,
120, 220, 240 VAC 50/60Hz (select voltage when ordering) |
|
 |
| Analog
Outputs |
Analog
Outputs 1,2,3
(dedicated monitor outputs
for each channel) |
Configuration |
Real-time
analog output voltage |
| Output Voltage |
± 3V = ±
FS on selected range |
| Accuracy |
Probe dependent |
| Frequency Response |
DC to 400 Hz |
Analog Output 4
(corrected output) |
Configuration |
Output voltage from
DAC |
| Output Voltage |
± 3V (Model
460-10: ± 10V) |
| Accuracy |
± 0.1% of
FS |
| Update Rate |
(same as field measurement)
|
| Connectors |
4 x
BNC jacks |
|
 |
| Physical |
| Net Weight |
16.5
lbs (7.5 kg) |
| Size (W x H x D) |
17.1" x 3.5"
x 14.5" (434mm x 89mm x 369mm) |
| Package Contents |
Gaussmeter, manual,
zero-gauss chamber, line cord
(probes must be purchased separately) |
| Software |
Labview driver (available